ZNB-XA | |
Pumping Mechanism | Curvilinear peristaltic |
IV Sets | Compatible with IV sets of any standard |
Flow Rate | 1~1100 ml/h in 0.1 ml/h increments |
Purge, Bolus | Purge when pump stops, bolus when pump starts, speed at 700 ml/h |
Accuracy | ±3% |
Volume | 1~9999 ml in 1ml increments |
Infusion Mode | ml/h, drop/min |
KVO Rate | 4 ml/h |
Alarms | Occlusion, air-in-line, door open, end program, low pressure, drop sensor, unattended pump after startup |
Additional Features | Real-time shift totals, automatic power switching, mute button, system malfunction |
Occlusion Sensitivity | 0.1-0.14 Mpa, adjustable |
Air-in-Line Detection | Ultrasonic detector |
Infusion Management System (nurse call) | Optional |
Drop Sensor | Optional |
Vehicle Power | Available |
Power Supply | AC: 110/230 VAC, 50/60 Hz, 20 VA
Battery: 9.6V±1.6V, rechargeable Vehicle power: 12V±1.2V |
Battery Life | 3 hours at the 30 ml/h |
Working Temperature | 10-30 |
Relative Humidity | 30-75% |
Atmospheric Pressure | 860-1060 hpa |
Size | 130*145*228 mm |
Weight | 2.5 kg |
Safety Classification | Class Ⅰ, type CF |